This site uses cookies to deliver our services and provide a basic level of website functionality. The four different types of cookies are: Strictly Necessary cookies, Functionality cookies, Performance cookies and Advertising cookies. By checking the box and clicking the button below, you are informing Rapid Electronics Ltd that you agree to the use of these cookies as detailed in our Privacy Policy.

Industry Schools FE/HE Site: Industry Site: Schools Site: FE/HE Site: Mobile

GW Instek GLC-9000 Leakage Current Tester

Order Code:  

90-6227

 
Brand: GW Instek  MPN: GLC-9000  EAN-13: 4719692504555
 Report an error
Qty Unit Price (Ex VAT)
1+ £2050.00

Get a quote for large quantities here

Units
Price per unit {{ IncVat ? 'Inc' : 'Ex' }} VAT
Additional quantity lead time 7 daysContact us for lead times on larger quantities
Out of StockStandard factory lead time 7 days

The GW Instek GLC-9000 leakage current tester, is used to perform leakage current (or called touch current) tests on medical electrical (IEC 60601-1) and general purpose electric and electronic (IEC 60990) equipment.

This tester engages with nine measurement networks (or called measuring device) to provide the simulation of human body while the EUT (equipment under test) is taking a leakage current testing in compliance with the specific standards or regulations such as IEC, UL, JIS...etc..

In order to provide a simple operation environment, the GW Instek GLC-9000 is equipped with a large TFT LCD touch panel to configure system as well as to present the measurement settings information and result simultaneously. Besides, there are 50 preset testing conditions, which conform to IEC60990 and other standards, for general electric equipment that can be recalled to reduce the setting time. In addition, 30 sets of empty memory are available for user defined setting.


  • Touch panel with colour LCD display
  • 9 Different measurement network to simulate the resistance of human body
  • 8 Different types of leakage current for medical and general electrical equipment
  • Various leakage current measuring mode: DC/AC/AC+DC/AC peak
  • Upper and lower limitation for pass/fail judgment
  • Meter mode with SELV/CONV function
  • 50 Preset test conditions conform to IEC 60990; 30 sets memories for customer defined
  • Various standard interfaces: RS232/GPIB/USB host and device/EXT I/O
  • Manufacturers part no. GW Instek GLC-9000

GeneralDisplay320 x 240 colour TFT LCD
OperationTouch panel
Measurement Device9 kinds
DCRanges50uA/500uA/5mA/25mA
Resolution0.01uA/0.1uA/1uA/10uA
Accuracy±1.0% fs/± (0.2% rdg+3dgt)
AC/AC+DCRanges50uA/500uA/5mA/25mA
Resolution0.01uA/0.1uA/1uA/10uA
Accuracy±2.0% fs/±(2.0% rdg+6dgt)- 10Hz
AC peakRanges500uA/1mA/10mA/75mA
Resolution0.1uA/1uA/10uA/100uA
Accuracy±4.0%fs/±2.5%fs/± (2.0% rdg+6dgt)
Meter ModeAC/DC/AC+DC10.0 to 300V
AC peak15.0 to 430V
SELV functionAvailable
EUTVoltage range85V to 300V
Current range0.5A to 10A
MemoryEmpty sets30 sets
Preset sets50 sets (conform to IEC 60990)
InterfacesFront panelUSB host
Rear panelRS-232C/GPIB/EXT I/O/USB device
Power SourceFor GLC-9000AC 100V/120V/220V/230V±10%, 50/60Hz
For EUTAC 85V to 250V, 50/60Hz
Dimensions330 (W) x 150 (H) x 350 (D) mm
WeightApprox. 5kg